Sd. Kolev et al., Mathematical modelling of potentiometric stripping analysis. Chemical stripping in quiet solutions, ANALYT CHIM, 377(1), 1998, pp. 13-19
A mathematical model of batch potentiometric stripping analysis is proposed
. It describes the case when the plating of the Hg electrode and the electr
odeposition of the analyte take place in a mechanically stirred measuring c
ell while the chemical stripping step takes place in the same solution but
without stirring. Between the electrodeposition step and the chemical strip
ping steps there is a period during which the moving solution is allowed to
settle completely. Potentiometric stripping experiments with standard solu
tions of Pb(II) in 0.1 M hydrochloric acid were performed at 15 and 60 s se
ttling times. The good agreement between the experimental results and the p
redictions of the model confirm its validity and show that it can be used s
uccessfully for better understanding and optimisation of batch potentiometr
ic stripping systems where the chemical stripping takes place without mecha
nical stirring. Depending on the requirements for sample throughput and sen
sitivity of a particular analysis, the appropriate size of the MF electrode
and the duration of the settling time can be easily selected using model s
imulations. (C) 1998 Published by Elsevier Science B.V.