Investigation of the wavelength accuracy of Brewer spectrophotometers

Citation
J. Grobner et al., Investigation of the wavelength accuracy of Brewer spectrophotometers, APPL OPTICS, 37(36), 1998, pp. 8352-8360
Citations number
7
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science","Optics & Acoustics
Journal title
APPLIED OPTICS
ISSN journal
00036935 → ACNP
Volume
37
Issue
36
Year of publication
1998
Pages
8352 - 8360
Database
ISI
SICI code
0003-6935(199812)37:36<8352:IOTWAO>2.0.ZU;2-B
Abstract
The wavelength accuracy of the Brewer spectrophotometer in the 280-360-nm s pectral range is improved by a new grating-drive mechanism and a new disper sion function derived from the Brewer geometry. With the new mechanism, the reproducibility of wavelength settings for spectral emission lines is bett er than 0.3 pm (0.0003 nm) and, with a thermally compensated version, the e ffect of temperature is less than 0.4 pm K-1. The new dispersion function f its spectral line positions better than the standard function and is less p rone to extrapolation error. Applying the new function to data from four ne w and ten standard drives shows that the new drives perform as well as the best of the standard ones and much better than the majority. (C) 1998 Optic al Society of America.