A. Del Guerra et al., YAP-PET: first results of a small animal Positron Emission Tomograph basedon YAP : Ce finger crystals, IEEE NUCL S, 45(6), 1998, pp. 3105-3108
We have built a small animal Positron Emission Tomograph (YAP-PET) based on
four matrices of 400 YAP:Ce finger crystals (2x2x30 mm(3) each) coupled to
Position Sensitive PhotoMultiplier Hamamatsu R2486-06. The system consists
of four detectors on a rotating gantry with a variable diameter from 10 an
d 25 cm. The read-out and data acquisition system are handled by NIM-CAMAC
standard electronics. The tomograph has a total Field Of View (FOV) of 4x4x
4 cm(3) that is appropriate for mice and rat studies. A spatial resolution
less than or equal to 2 mm over 3x3x3 cm(3) FOV has been measured, with a p
eak sensitivity for a point source of about 614 cps/mu Ci for a gantry diam
eter of 15 cm and an energy The system consists of four detectors on a rota
ting gantry with threshold of 50 keV.