Mg. Salmas et al., Estimation of artifacts induced by the Ilizarov device in quantitative computed tomographic analysis of tibiae, INJURY, 29(9), 1998, pp. 711-716
Citations number
6
Categorie Soggetti
Aneshtesia & Intensive Care
Journal title
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED
This study concerns the estimation of the artifacts introduced by the metal
lic frame of an Ilizarov apparatus during quantitative computed tomography
for the assessment of neoosteogenesis. Ten cadaver tibiae were tomographed
before and after the mounting of an Ilizarov device. The increase in mean d
ensity and in the relative number of pixels were used as indicators of the
artifacts induced by the apparatus. No significant influence by the device
was recorded in regions corresponding to a corticotomy site. In regions clo
se to the metallic rings of the Ilizarov apparatus, measurements are distor
ted significantly, but with no influence on the assessment of neoosteogenes
is. (C) 1998 Published by Elsevier Science Ltd. All rights reserved.