The dose distribution for an ovoid with a new tungsten shielding design was
determined using Monte Carlo simulation. Standard Cesium-137 tube sources,
tungsten shielding, and aluminum ovoid applicator were each modeled as a c
ollection of solid objects. Dose was calculated in planes above, below, in
front of, and on the sides of the colpostat. The Monte Carlo results were c
ompared with the results from a parametrized calculation algorithm and good
agreement was obtained. The dose distribution matrix derived from the para
metrized algorithm can be used for clinical treatment planning. (C) 1998 Am
erican Association of Physicists in Medicine. [S0094-2405(98)01412-6].