Yl. Feng et al., SEMIAUTOMATIC DETERMINATION OF TIN IN MARINE MATERIALS BY CONTINUOUS-FLOW HYDRIDE GENERATION INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY, Fresenius' journal of analytical chemistry, 357(7), 1997, pp. 822-826
A semi-automated continuous flow hydride generation system with induct
ively coupled plasma atomic emission spectrometry (ICP-AES) was used f
or the determination of tin in marine materials. The effects of acids
(H2SO4 and HCl) were studied. The analytical parameters were thoroughl
y investigated. Under optimized conditions, the detection limit is 0.4
ng/ml. Interferences from transition elements were investigated and s
even masking reagents were tested. L-cysteine hydrochloride monohydrat
e (1%) was used to mask the interferences from foreign ions. Finally,
the accuracy, checked with a marine standard reference material obtain
ed from the National Research Council (NRC), was within the certified
value.