PURPOSE: To measure directly the grid performance of mammography units for
the range of breast thicknesses and x-ray tube potentials encountered in cl
inical practice.
MATERIALS AND METHODS: Contrast improvement factors and Bucky factors were
determined for four mammographic units as a function of x-ray tube potentia
l (25, 30, and 35 kVp), phantom thickness (2, 4, and 8 cm) and, on one unit
, three target-filter combinations. Three units used a linear grid; one, a
cellular grid. Two methods were used for nongrid measurements.
RESULTS: For all units tested, contrast improvement factor increased with i
ncreased phantom thickness and with increased kilovolt peak level for the 8
-cm-thick phantom and changed little with kilovolt peak level for 2- and 4-
cm-thick phantoms. At 25 and 30 kVp, contrast improvement factor performanc
e with the linear grids was comparable; with the cellular grid, it was 5%-1
0% higher. In all cases, the Bucky factor increased with increased phantom
thickness and decreased with increased tube potential.
CONCLUSION: Differences in grid performance exist. At 25 and 30 kVp, the ce
llular grid exhibited superior contrast improvement factor performance, whe
reas one of the linear grids exhibited superior Bucky factor performance. M
easured contrast improvement and Bucky factors are dependent on nongrid tec
hnique. Cassette tunnels introduce scatter and should not be used with nong
rid or-magnification techniques.