Je. Tanner et al., DOSE-EQUIVALENT RESPONSE OF PERSONAL NEUTRON DOSIMETERS AS A FUNCTIONOF ANGLE, Radiation protection dosimetry, 70(1-4), 1997, pp. 165-168
Citations number
8
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
The measured and calculated dose equivalent response as a function of
angle has been examined for an albedo-type thermoluminescence dosemete
r (TLD) that was exposed to unmoderated and D2O-moderated Cf-252 neutr
on sources while mounted on a 40 x 40 x 15 cm(3) polymethylmethacrylat
e phantom. The dosemeter used in this study is similar to many neutron
personal dosemeters currently in use. The detailed construction of th
e dosemeter was modelled, and the dose equivalent response was calcula
ted, using the MCNP code. Good agreement was found between the measure
d and calculated values of the relative dose equivalent angular respon
de for the TLD albedo dosemeter. The relative dose equivalent angular
response was also compared with the values of directional and personal
dose equivalent as a function of angle published by Siebert and Schuh
macher.