DETERMINATION OF NEUTRON ROOM SCATTERING CORRECTIONS IN RCLS CALIBRATION FACILITY AT KAERI

Citation
Sc. Yoon et al., DETERMINATION OF NEUTRON ROOM SCATTERING CORRECTIONS IN RCLS CALIBRATION FACILITY AT KAERI, Radiation protection dosimetry, 70(1-4), 1997, pp. 349-351
Citations number
10
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
70
Issue
1-4
Year of publication
1997
Pages
349 - 351
Database
ISI
SICI code
0144-8420(1997)70:1-4<349:DONRSC>2.0.ZU;2-O
Abstract
Neutron room scattering corrections that should be made when neutron d etectors are calibrated with a D2O moderated Cf-252 neutron source in the centre of a calibration room are considered. Such room scattering corrections are dependent on specific neutron source type, detector ty pe, calibration distance, and calibration room configuration. Room sca ttering corrections for the responses of thermoluminescence dosemeter and two different types of spherical detectors to a neutron source in the Radiation Calibration Laboratory (RCL) neutron calibration facilit y at the Korea Atomic Energy Research Institute (KAERI) were experimen tally determined and are presented. The measured room scattering resul ts are then compared with theoretical results calculated by predicting room scattering effects in terms of parameters related to the specifi c configuration. Agreement between measured and calculated scattering correction is generally about 10% for three kinds of detectors in the calibration facility.