M. Yudelev et al., EVALUATION OF THE DOSE ALGORITHM-BASED ON THE SCATTER MODEL APPLIED FOR NEUTRON THERAPY TREATMENT PLANNING, Radiation protection dosimetry, 70(1-4), 1997, pp. 513-516
Citations number
6
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
The algorithm based on the scatter model was successfully applied in t
otal dose (neutron plus gamma) calculations for Harper Hospital d(48.5
) + Be neutron beam. The central axis and the lateral beam profiles as
well as the peak scatter factors were measured for a number of rectan
gular and irregularly shaped fields. These data were used to calculate
the tissue- and scatter-to-air ratios (TAR and SAR) as well as to det
ermine the beam parameters pertinent to the treatment planning algorit
hm and to evaluate its accuracy in clinical cases. The agreement betwe
en the measured data and the data derived from the treatment planning
program is usually within 2% for the central axis points and about 5%
for the lateral beam profiles.