R. Accinni et al., High-performance liquid chromatographic determination of total plasma homocysteine with or without internal standards, J CHROMAT A, 828(1-2), 1998, pp. 397-400
Hyperhomocysteinemia is an independent risk factor for atherosclerosis and
vascular occlusive disease. Assessment of total plasma concentration of hom
ocysteine (tHcys) requires accurate and reproducible measurements. The aim
of this study was to test a rapid isocratic HPLC method for tHcys analysis
with an internal standard (I.S.) of alpha-mercaptopropionylglycine (MPG), 2
-mercaptoethylamine (ME), or N-acetylcysteine (NAC) or without I.S., and to
verify whether the use of an I.S. improves the precision. The method witho
ut I.S. showed an excellent linearity (y=1.59x-0.15, r=1), recovery (100%)
and a within-assay relative standard deviation (R.S.D.) of 1.2%. Instead, i
n our hands, the presence of I.S,s decreased the reproducibility (within-as
say R.S.D. ranged from 3.5 to 6.5%) and lengthened the chromatogram by up t
o four to five times. In conclusion, HPLC measurement of plasma tHcys witho
ut I.S. improves accuracy with respect to determination with I.S.; moreover
, this approach allows to routinely process larger amounts of plasma sample
s. (C) 1998 Published by Elsevier Science B.V. All rights reserved.