Patch migration: A serious late complication of thoracotomy lead systems

Citation
I. Singer et al., Patch migration: A serious late complication of thoracotomy lead systems, PACE, 21(12), 1998, pp. 2616-2620
Citations number
18
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
21
Issue
12
Year of publication
1998
Pages
2616 - 2620
Database
ISI
SICI code
0147-8389(199812)21:12<2616:PMASLC>2.0.ZU;2-0
Abstract
Thoracotomy patch leads used for implantable cardioverter defibrillators (I CDs) are generally safe and effective. We describe two patients in whom a l ate complication of patch lead migration occurred years after the original implants, causing a bronchopleural fistula in one and lingular lobe collaps e in the other patient. We conclude that patch migration is a late but poss ible complication of extrapericardial ICD leads, and should be suspected in patients who present with hemoptysis, atypical pneumonia, or lung collapse after the initial ICD surgery.