Twiddler's syndrome variants in ICD patients

Citation
Ng. Boyle et al., Twiddler's syndrome variants in ICD patients, PACE, 21(12), 1998, pp. 2685-2687
Citations number
5
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
21
Issue
12
Year of publication
1998
Pages
2685 - 2687
Database
ISI
SICI code
0147-8389(199812)21:12<2685:TSVIIP>2.0.ZU;2-5
Abstract
Three cases of twiddler's syndrome in patients nifh an abdominally implante d cardioverter defibrillator (ICD) and an endocardial lead system are repor ted. The condition was detected when an increase in pacing threshold or lea d impedance was noted at routine follow-up. Successful revision as possible in all patients; however, a recurrence in one patient necessitated removal of the device. Elderly female patients with some degree of obesity appear most susceptible. Minimizing pocket size and suturing of the der ice to the fascia should decrease the chance of developing this problem.