P. Dieuzaide et al., A complication of pacemaker lead extraction: Paradoxical embolism of a lead fragment in a leg artery, PACE, 21(12), 1998, pp. 2699-2700
Citations number
5
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Extraction of three chronically implanted pacemaker leads was performed via
the implant vein in a 71-year-old man. One of the leads broke in the subcl
avian scar tissue. The lead fragment migrated into the left peroneotibial t
runk artery. Transesophageal echocardiography showed patent foremen or ale
associated with right-to-left atrial shunt. This complication of lead extra
ction is discussed along with the role of echocardiography prior to lead re
moval.