A method of reducing oxygen induced drift in iridium oxide pH sensors

Citation
J. Hendrikse et al., A method of reducing oxygen induced drift in iridium oxide pH sensors, SENS ACTU-B, 53(1-2), 1998, pp. 97-103
Citations number
24
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
SENSORS AND ACTUATORS B-CHEMICAL
ISSN journal
09254005 → ACNP
Volume
53
Issue
1-2
Year of publication
1998
Pages
97 - 103
Database
ISI
SICI code
0925-4005(19981126)53:1-2<97:AMOROI>2.0.ZU;2-G
Abstract
A novel way of reducing drift problems of the electrode potential of metal oxide pH sensors is presented. The method employs a FET-structure that uses the conducting metal oxide electrode as a gate contact. From the threshold voltage of this structure a signal can be derived which is independent of the electrode potential and can be used to compensate for the drift of it. In addition to the reduced drift, the compensation leads to an almost ideal nernstian response. First a theoretical explanation is given, which is the n confirmed by measurement results. (C) 1998 Elsevier Science S.A. All righ ts reserved.