C. Galan-vidal et al., Glucose biosensor strip in a three electrode configuration based on composite and biocomposite materials applied by planar thick film technology, SENS ACTU-B, 52(3), 1998, pp. 257-263
An amperometric glucose biosensor strip based on a three electrodes (counte
r, reference and working) planar configuration was fabricated using thick f
ilm technology. Glucose oxidase (GOD) was immobilized in the bulk of a grap
hite-epoxy composite paste. This paste was used to screen-print the working
electrode on a glass fibre board. The integration on the same support of p
seudo-reference electrode was obtained by screen printing a commercial silv
er-epoxy paste, and subsequent electrochemical chlorinization in KCl 1 M at
1 + V (SCE). Finally, the counter electrode was integrated using the same
technique by depositing a graphite-epoxy composite paste, taking into accou
nt the counter/working electrode area ratio. Response characteristics of a
glucose biosensor strip based on this transducer are similar to those of pr
eviously reported one electrode configuration glucose biosensor constructed
using the same biocomposite material [Galan-Vidal and Munoz, Sens. Actuato
rs B 45 (1997) 55-62]. (C) 1998 Elsevier Science S.A. All rights reserved.