I. Isildak et al., Spectrophotometric determination of copper(II) at low mu g 1(-1) levels using cation-exchange microcolumn in flow-injection, TALANTA, 48(1), 1999, pp. 219-224
A simple spectrophotometric flow-injection method is reported for the highl
y sensitive and fast determination of copper(II). The method is based on th
e formation of coloured Cu(II)-(4-methylpiperidinedithiocarbamate)(2) compl
ex when the copper solutions are introduced into a tertiary reagent stream
containing 4-methylpiperidinedithiocarbamate. The coloured complex is then
selectively monitored at 435 nm. To increase interactions between copper(II
) and colour forming reagent and preconcentrate of copper(II), a microcolum
n containing strong cation exchange resins was placed between injection man
ifold and spectrophotometer. The system required no mixing chamber and allo
wed a sample throughput > 60 sample h(-1). The calibration graph was linear
in the range 5-100 mu g 1(-1). The detection limit was < 0.5 mu g 1(-1) fo
r 20 mu 1 injection volume of copper(II) ion solution. The developed method
was applied to environmental, copper processing water, and ore samples. (C
) 1999 Elsevier Science B.V. All rights reserved.