Heparin-induced coagulation disturbance through machine autotransfusion

Citation
M. Vorweg et al., Heparin-induced coagulation disturbance through machine autotransfusion, ANAESTHESIS, 47(12), 1998, pp. 979-981
Citations number
10
Categorie Soggetti
Aneshtesia & Intensive Care
Journal title
ANAESTHESIST
ISSN journal
00032417 → ACNP
Volume
47
Issue
12
Year of publication
1998
Pages
979 - 981
Database
ISI
SICI code
0003-2417(199812)47:12<979:HCDTMA>2.0.ZU;2-X
Abstract
Disorders in blood coagulation during the use of autotransfusion have been reported in recent literature. We wondered whether or not heparine, remaini ng in the prepared sam pie of retransfusion blood might be responsible for these disturbances. Methods: Therefore we created a setting in which heparine was added to anim al blood in order to imitate clinical situations during the use of the CELL SAVER. Results: According to our results the blood shows no irregular heparine loa d as long as the machine is used following the operating instructions. But if the volume of rinsing liquid is decreased or the pumpflow is increased, the heparine load is increased enormously (Table 1). Conclusions: If the CELL SAVER is used with reduced volume of rinsing liqui d or higher velocity of the pump, the coagulation status of the patient has to be monitored meticously.