Whole Blood Capcellia CD4/CD8 immunoassay for enumeration of CD4+ and CD8+peripheral T lymphocytes

Citation
D. Carriere et al., Whole Blood Capcellia CD4/CD8 immunoassay for enumeration of CD4+ and CD8+peripheral T lymphocytes, CLIN CHEM, 45(1), 1999, pp. 92-97
Citations number
16
Categorie Soggetti
Medical Research Diagnosis & Treatment
Journal title
CLINICAL CHEMISTRY
ISSN journal
00099147 → ACNP
Volume
45
Issue
1
Year of publication
1999
Pages
92 - 97
Database
ISI
SICI code
0009-9147(199901)45:1<92:WBCCIF>2.0.ZU;2-0
Abstract
We evaluated the Whole Blood Capcellia(R) CD4/CD8, an immunoenzymatic metho d that provides absolute counts of CD4+ and CD8+ T cells in peripheral bloo d. The assay is based on the separation of T cells by use of an anti-CD2 ma gnetic bead suspension, followed by reaction of the CD4 or CD8 molecules wi th the corresponding monoclonal antibody coupled to peroxidase. CD4-positiv e monocytes were excluded from the assay. Freeze-dried magnetic bead-T-cell complexes were used as calibrators. Capcellia counts from HIV-1-infected p atients were compared with those obtained by flow cytometry as the comparis on method. The results by Capcellia correlated well with those by flow cyto metric analysis: r(2) = 0.95; P <0.001; (y = 0.96x - 22.1); S-y/x = 64 for CD4; r(2) = 0.81; P <0.001; (y = 1.26x - 76.4); S-y/x = 139 for CD8; n = 76 . The correlation between CD4+ T-cell counts determined by two trained expe rimenters was significant (r(2) = 0.96). Our results indicate that this new ELISA technique far lymphocyte immunophenotyping is an efficient alternati ve to flow cytometry.