Jj. Prete et al., THE COUPLING OF ANISOTROPY AND RADIAL DOSE FUNCTIONS FOR PD-103 AND I-125 FOR USE WITH A COMMERCIAL TREATMENT PLANNING SYSTEM, Medical physics, 24(4), 1997, pp. 523-525
Many commercial treatment planning systems available today employ trad
itional dose calculation formulae in their interstitial brachytherapy
source calculation algorithms. The 1995 AAPM report on interstitial br
achytherapy source dosimetry recommended a new dose calculational form
alism and presented a technique for adopting it on systems which embod
y traditional formalism. In order to comply with these recommendations
on our system, the transformations for implementing a one dimensional
isotropic point source model were modified by coupling the published
anisotropy and radial dose corrections and fitting them to a fifth ord
er polynomial. Using this approach, a more accurate dose calculation i
s obtained.