PROCESSOR QUALITY-CONTROL IN LASER IMAGING-SYSTEMS

Citation
Tm. Bogucki et al., PROCESSOR QUALITY-CONTROL IN LASER IMAGING-SYSTEMS, Medical physics, 24(4), 1997, pp. 581-584
Citations number
5
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
00942405
Volume
24
Issue
4
Year of publication
1997
Pages
581 - 584
Database
ISI
SICI code
0094-2405(1997)24:4<581:PQILI>2.0.ZU;2-6
Abstract
Sensitometric techniques for performing processor quality control in l aser imaging systems are analyzed in this study. The sensitivity of co nventional x-ray films using simulated screen-light sensitometry is co mpared with helium-neon (HeNe) laser film exposed with a simulated red -light sensitometer, a standalone (reference) laser sensitometer, an e xperimental (unstable) laser sensitometer, and laser printers. Infrare d (IR) laser film exposed with an IR laser diode and a simulated IR se nsitometer are also evaluated. It is demonstrated that laser-generated step tablets provide an easy and reliable method of performing laser film processor quality control. (C) 1997 American Association of Physi cists in Medicine.