Z. Vujic et al., DENSITOMETRIC DETERMINATION OF METOPROLOL TARTRATE IN PHARMACEUTICAL DOSAGE FORMS, Journal of pharmaceutical and biomedical analysis, 15(5), 1997, pp. 581-585
This paper describes a simple densitometric method for the determinati
on of metroprolol tartrate in tablets and ampoules. After separation o
n silica gel GF254 plates, using acetone-methanol-triethylamine as the
mobile phase for the tablets and acetone-triethylamine for ampoules,
the chromatographic zones corresponding to the spots of metoprolol wer
e scanned. Quantitation was performed using a computer-controlled Cama
g TLC scanner and applying five-point calibration with polynomial regr
ession. The calibration function was established in the ranges 1-28 mu
g for tablets and 1-9 mu g for ampoules. The results obtained are pre
cise and reproducible, with recovery values of 99.1-99.4%.