Combination of an intensified charge coupled device with an echelle spectrograph for analysis of colloidal material by laser-induced plasma spectroscopy
C. Haisch et al., Combination of an intensified charge coupled device with an echelle spectrograph for analysis of colloidal material by laser-induced plasma spectroscopy, SPECT ACT B, 53(12), 1998, pp. 1657-1667
An evaluation of the adaptation of an echelle spectrograph to an intensifie
d charge coupled device (ICCD) for Laser-induced Plasma Spectroscopy (LIPS)
analysis of environmentally relevant heavy metal hydrocolloids deposited o
n a membrane filter is presented. For artificial and natural colloids we co
uld demonstrate that an echelle spectrograph with an ICCD allows a spectral
resolution and enlarged spectral coverage superior to a conventional LIPS
detector system based on a Czerny-Turner spectrograph and an intensified di
ode-array. Especially natural samples were not accessible with the conventi
onal system, while the echelle system permitted an unequivocal multivariate
identification of a set of emission lines from a large set of individual e
lements. For different heavy metal hydrocolloids we could verify that no si
gnificant difference in terms of limit of detection exists between the eche
lle/ICCD system and the conventional system. In addition, we illustrated fo
r Al2O3 colloids the utilization of the echelle spectrograph for plasma dia
gnostics, which can be important for normalization in case of variable plas
ma conditions and self-absorption. (C) 1998 Elsevier Science B.V. All right
s reserved.