A simplified extracorporeal circulation (ECC) assemblage with autogenous ox
ygenation (AO) using a single centrifugal pump was tested in dogs. The tran
spulmonary gradient was obtained by increasing pressure in the right atrium
through volume expansion and decreasing it in the left atrium by collectin
g the blood from the pulmonary veins in a reservoir placed below the level
of the animal, generating a siphon effect. This arrangement dispenses with
a right side pump. The heart was electrically fibrillated after perfusion w
as started and defibrillated at the end of the bypass. This ECC circuit all
owed the maintenance of adequate hemodynamic and blood gas parameters durin
g the bypass. The operating field and the mobility of the heart were equiva
lent to that of conventional cardiopulmonary bypass (CPB). We conclude that
the use of a single centrifugal Dump simplifies the autogenous oxygenation
approach, making it a practical choice for the coronary artery bypass graf
t (CABG) procedure.