Simple procedure to eliminate concentrated sodium bicarbonate and carbonate with reusable strong cation-exchange cartridges for ion chromatographic analysis of formate and oxalate
Zl. Cai et Y. Katsumura, Simple procedure to eliminate concentrated sodium bicarbonate and carbonate with reusable strong cation-exchange cartridges for ion chromatographic analysis of formate and oxalate, J CHROMAT A, 829(1-2), 1998, pp. 407-410
A straightforward method with strong cation-exchange cartridges to eliminat
e bicarbonate and carbonate interference in IC analysis of trace formate an
d oxalate is described. The recoveries and detection limits for formate and
oxalate were 101+/-4% and 1.6 mu M, 87+/-2% and 0.8 mu M, respectively, an
d independent of individual cartridges, pH and concentrations of interferin
g ions in the matrix. The cartridges were regenerated with 100 mM Na2CO3, 5
0 mM H2SO4 and water, and reused for at least 50 times. This method was app
lied to quantification of radiolysis products of gamma-ray-irradiated bicar
bonate and carbonate aqueous solutions with ion chromatography. (C) 1998 El
sevier Science B.V. All rights reserved.