We investigated the performances of HbA1c determination by a latex enhanced
turbidimetric immunoassay using the specific monoclonal antibodies (Unimat
e, Roche) against the beta-N-terminal fragments. The coefficients of variat
ion ranges from 1.7 to 3.8% within assay (n = 30) and from 3.9 to 4.9% betw
een assay (n = 20).
The assay was linear from 2.5 to 14.9% of HbA1c. No interferences was found
from fetal, carbamylated, or variant (S) hemoglobins and from labile Schif
f adduct with glucose. The following relationship was derived from fresh sa
mple comparison between HPLC (Diamat-BioRad) (x) and immunoassay (y) method
: y = 0.971 x + 0.87%, r = 0.98, n = 115. The immunoassay provides a highly
precise and specific method for HbA1c. (C) 1999 Wiley-Liss, Inc.