Pl. Bailey et al., Evaluation in volunteers of the VIA V-ABG automated bedside blood gas, chemistry and hematocrit monitor, J CLIN M C, 14(5), 1998, pp. 339-346
Objective. To evaluate the VIA V-ABG (VIA Medical Corp.) point-of-care bloo
d gas and chemistry monitor in healthy human volunteers, with particular em
phasis on the measurement of blood gases. Methods. Experimental conditions
were varied by intermittently subjecting volunteers to either isocapnic hyp
ercapnia (end-tidal (ET), PETCO2 = 50 +/- 2 mmHg, ETPO2 = 130 +/- 5 mmHg) o
r isocapnic hypoxia (PETCO2 = 42 +/- 2, PETO2 + 45 +/- 2 mmHg) in addition
to room air breathing. Measurements by the VIA V-ABG device were compared w
ith paired samples and measurements performed by two ABL Radiometers (505 a
nd 500). Analysis of results includes bias and precision plots and comparis
on of results with minimal performance criteria as established by CLIA. Res
ults. Nineteen volunteers yielded 222 matched samples. The range of values
were 7.32-7.61 for pH, 20.9-51.6 mmHg for PCO2, 27.9-184.5 mmHg for PO2, 13
1-141 mmol/l for Na, 3.1-4.1 mmol/l for K, and 30.0-50.4% for hematocrit. B
ias and precision (+/- 2 sd) for pH was 0.01 and 0.03, for PCO2 was 0.4 and
4.8, for PO2 was 1.0 and 17.0, for Na was -0.3 and 5.2, for K was 0.1 and
0.2, and for Her was 2.0 and 5.4. Conclusions. Over the range of blood gas
values assessed, blood gas measurements by the VIA V-ABC device were clinic
ally acceptable and met minimal performance criteria utilizing current Medi
care CLIA proficiency standards. Performance criteria were also met by the
VIA V-ABG device for Na, K, and Hct measurements but the range of values wa
s too narrow to allow characterization of clinical acceptability. The VIA V
-ABG device appears to perform well compared with the results which have be
en published for other point-of-care devices. Comparison between different
studies investigating point-of-care devices is difficult due to several fac
tors (range of values measured, comparison device, population studied, etc.
). Some of these instruments, including the VIA V-ABG device, may serve qui
ts well as point-oi-care devices to perform certain tests at the bedside. W
hether or not any of these devices can substitute for traditional laborator
y blood gas and chemistry measurements remains an issue that is not adequat
ely studied.