Region of interest (ROI) imaging has previously been proposed as a means of
reducing x-ray fluoroscopy radiation dose. Previous ROI attenuators made o
f partially attenuating metal plates change beam quality, which may lead to
uncertainty in image restoration procedures. The design and construction o
f a prototype moving segments ROI attenuator (MS-ROI), which maintains beam
quality across the whole field of view is described. The x-ray beam intens
ity is halved by 36 lead segments which are rapidly rotated between the x-r
ay tube and patient, with a central hole projecting a circular ROI at full
intensity. Image processing techniques with automatic detection of the ROI
boundary were used to homogenize image brightness across the whole image. I
mages restored using these techniques were judged to be visually acceptable
, with a good match between pixel values inside and outside the ROI. Image
contrast within the ROI was improved by 18% due to reduced scatter and veil
ing glare from the periphery. The introduction of the MS-ROI attenuator als
o results in a 48% increase in statistical noise in the area outside the RO
I, with no significant change in object contrast. The patient entrance dose
measured using the dose area product (DAP) method was reduced by 53.4% und
er manual exposure control, with the dose to operators reduced by 48.4% und
er automatic brightness control. Further work is needed to determine whethe
r the attenuator can be used with pulsed fluoroscopy, and to reduce vibrati
onal effects on the ROI boundary. The MS-ROI attenuator provides a more con
stant ratio of central-to-peripheral image intensity, and maintains uniform
beam quality and image contrast across the whole image in comparison to si
mple metal plate attenuators. (C) 1999 American Association of Physicists i
n Medicine. [S0094-2405(99)00601-X].