Design for an accelerator-based orthogonal epithermal neutron beam for boron neutron capture therapy

Citation
Da. Allen et al., Design for an accelerator-based orthogonal epithermal neutron beam for boron neutron capture therapy, MED PHYS, 26(1), 1999, pp. 71-76
Citations number
14
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MEDICAL PHYSICS
ISSN journal
00942405 → ACNP
Volume
26
Issue
1
Year of publication
1999
Pages
71 - 76
Database
ISI
SICI code
0094-2405(199901)26:1<71:DFAAOE>2.0.ZU;2-4
Abstract
This paper is concerned with the proposed Birmingham accelerator-based epit hermal neutron beam for boron neutron capture therapy (BNCT). In particular , the option of producing a therapy beam at an orthogonal direction to the incoming protons is considered. Monte Carlo radiation transport simulations , both with and without a head phantom, have shown that an orthogonal beam geometry is not only acceptable but is indeed beneficial, in terms of a low er mean neutron energy and an enhanced therapeutic ratio for the same usefu l neutron fluence in the therapy beam. Typical treatment times for various beam options have been calculated, and range from 20 to 48 min with a 5 mA beam of 2.8 MeV protons, if the maximum photon-equivalent dose delivered to healthy tissue is to be 12.6 Gy Eq. The effects of proton beam diameter up on the therapy beam parameters have also been considered. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00701-4].