C. Simon et al., RADIOFREQUENCY HEATED BIPOLAR ENERGY IN A RTERIES - EXPERIMENTAL RESULTS WITH A NEW CATHETER, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 166(3), 1997, pp. 248-253
Purpose: To evaluate the effects of radiofrequency heated bipolar prob
es on the human arterial wall. Methods: Under standardised conditions,
lesions were set on the inner layer of a piece of human cadaver arter
ial wall, using a radiofrequency-heated bipolar catheter. These lesion
s were measured and described both macroscopically and microscopically
. Lesion size, applied energy levels and application conditions were c
orrelated. Results: The size of a lesion depended significantly on the
size of the catheter, angle and pressure of catheter application (p <
0.05) and length of impulse (p = 0.002). There was no influence from
the energy level employed. It proved impossible to treat calcified pla
ques. The depth of a lesion depended on the angle of catheter applicat
ion only No perforations were seen. Conclusions: Radiofrequency-heated
bipolar catheters ablate the tissue of the arterial wall. Increasing
tissue resistance resulting from higher energy levels prevents perfora
tion in the experimental set-up. Calcified plaques cannot be treated b
y this method.