Polarographic determination of cyanide as contaminant in pralidoxime mesylate formulations

Citation
Pc. Do Nascimento et al., Polarographic determination of cyanide as contaminant in pralidoxime mesylate formulations, TALANTA, 48(2), 1999, pp. 341-346
Citations number
33
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
TALANTA
ISSN journal
00399140 → ACNP
Volume
48
Issue
2
Year of publication
1999
Pages
341 - 346
Database
ISI
SICI code
0039-9140(199902)48:2<341:PDOCAC>2.0.ZU;2-D
Abstract
A method for the polarographic determination of cyanide as contaminant in p ralidoxime mesylate (PM) formulations was developed. The volatile cyanide f ormed in the formulations was stabilized as tetracyanonickelate (TCN) anion complex after reaction with ammoniacal NI(II) solution. The stable TCN ani on complex (K-stb = 10(31)) was determined by anodic stripping voltammetry sat the hanging mercury drop electrode (HMDE). The polarographic signal was proportional to the cyanide concentration and the high concentration of PM did not interfere. The linear range of calibration was from 1.2 to 16 mu g cyanide with r = 0.998. The RSD was 1.3% (n = 5) for 2.4 mu g cyanide and a detection limit of 0.8 mu g cyanide was calculated. The proposed method i s adequate as a quality control of PM formulations. (C) 1999 Elsevier Scien ce B.V. All rights reserved.