Diagnosis of phase shift in a temperature-modulated calorimetric method
Citation
I. Hatta et N. Katayama, Diagnosis of phase shift in a temperature-modulated calorimetric method, J THERM ANA, 54(2), 1998, pp. 577-584
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF THERMAL ANALYSIS AND CALORIMETRY
SICI code
1388-6150(1998)54:2<577:DOPSIA>2.0.ZU;2-#
Abstract
In a temperature-modulated calorimetric method using the same apparatus as
a standard differential scanning calorimeter, we have to pay attention to t
he thermophysical parameters of the apparatus, which cause phase shift in a
c temperatures, such as heat capacity of base plate, heat capacity of a pan
, thermal conductance between a heater and base plate, and thermal conducta
nce between a pan and base plate. We performed the analysis of the thermal
system oi the apparatus with these parameters. Beside the theoretical consi
deration, we carried out heat capacity measurement in a wide range of modul
ation periods. We found that the experimental results were well-expressed i
n terms of these thermophysical parameters.