R. Wens et al., A previously undescribed side effect of icodextrin: Overestimation of glycemia by glucose analyzer, PERIT DIA I, 18(6), 1998, pp. 603-609
Objective: Serious discrepancies between glycemia measurements obtained wit
h an Accutrend Sensor (Boehringer Mannheim GmbH, Mannheim, Germany) type an
alyzer (based on a glucose dehydrogenase enzymatic reaction) and measuremen
ts obtained in the laboratory by a reference method (hexokinase) have been
found in an insulin-requiring, diabetic, continuous ambulatory peritoneal d
ialysis (CAPD) patient treated with icodextrin 7.5% (Extraneal; Baxter Heal
thcare SA, Castlebar, Ireland), a new osmotic agent for peritoneal dialysis
. We therefore investigated the respective role of the Analyzer and of the
glucose polymer in this hitherto undescribed problem.
Design: Glycemia was measured simultaneously on venous blood using a refere
nce laboratory technique, and on capillary blood using the Accutrend Sensor
glucose analyzer in three groups of CAPD patients: 6 patients on Extraneal
for at least 1 week, 6 patients receiving their first Extraneal exchange,
and 8 patients never exposed to Extraneal. In the first group of patients,
glycemia was also measured with another analyzer (Glucocard; Menarini Diagn
ostics, Firenze, Italy) using a different enzymatic reaction (glucose oxida
se). In a separate study, whole blood of a normal subject was spiked with c
oncentrated solutions of glucose and icodextrin and some of its metabolites
(maltose, maltotriose, maltopentaose). Once again, comparative measurement
s of glycemia were performed with the Accutrend Sensor, with two other kits
using a glucose dehydrogenase enzyme reaction, and with the hexokinase ref
erence method.
Results: In 6 CAPD patients treated with once-daily exchanges with Extranea
l for a minimum of 7 consecutive days, we confirmed overestimation of glyce
mia by the Accutrend Sensor of 65 +/- 26 mg/dL compared to reference values
(p < 0.01), and of 69 +/- 25 mg/dL(p < 0.001) compared to measurements obt
ained with the Glucocard monitor. In 6 other CAPD patients studied at the e
nd of one single icodextrin exchange, overestimation of 61 +/- 11 mg/dL was
already present (p < 0.001). On the other hand, in 8 CAPD patients never t
reated with icodextrin, there was no discrepancy between the Accutrend Sens
or readings and reference values. The measurements in spiked blood confirme
d that only the Accutrend Sensor overestimates glycemia in the presence of
maltose and glucose polymers. The overestimation decreased as the molecular
size of the saccharides added to blood increased. There was no overestimat
ion when other kits using a dehydrogenase enzyme were tested.
Conclusion: The overestimation observed is probably related to the presence
of oligosaccharides (mainly maltose), derivatives of glucose polymers pres
ent in Extraneal and absorbed via the peritoneal route, in the blood of pat
ients treated with icodextrin. The glucose dehydrogenase characterizing the
Accutrend Sensor, an enzyme of the pyrroloquinolinequinone class, very lik
ely reacts with the free reducing group of the glucose molecule located at
the end of each saccharide chain. This would not be the case for the Glucoc
ard monitor using glucose oxidase, for other kits using glucose dehydrogena
se, and for the reference method based on hexokinase. The Accutrend Sensor
type of analyzers are therefore not suitable for regular monitoring of glyc
emia in diabetic PD patients treated with icodextrin.