Requirements for an active spectroscopy diagnostic with neutral beams on the RFX reversed field pinch

Citation
L. Carraro et al., Requirements for an active spectroscopy diagnostic with neutral beams on the RFX reversed field pinch, REV SCI INS, 70(1), 1999, pp. 861-864
Citations number
20
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
70
Issue
1
Year of publication
1999
Part
2
Pages
861 - 864
Database
ISI
SICI code
0034-6748(199901)70:1<861:RFAASD>2.0.ZU;2-6
Abstract
The conceptual feasibility of an active spectroscopic diagnostic on reverse d field experiment (RFX) to measure the radial profiles of ion temperature, toroidal and poloidal drift velocities, and main impurity densities is exa mined. In particular a charge exchange recombination spectroscopy experimen t is simulated for carbon and helium impurities. The motional Stark effect on the beam atoms injected into RFX is estimated as a means to measure the magnetic field pitch and intensity. A modulated beam injector (500 Hz) with 2 A of equivalent current and an energy of 50 keV/amu appears to be adequa te to perform the measurements over a reasonable range of operational regim es. (C) 1999 American Institute of Physics. [S0034-6748(99)70601-5].