A CONFORMATION NUMBER TO QUANTIFY THE DEGREE OF CONFORMALITY IN BRACHYTHERAPY AND EXTERNAL-BEAM IRRADIATION - APPLICATION TO THE PROSTATE

Citation
A. Vantriet et al., A CONFORMATION NUMBER TO QUANTIFY THE DEGREE OF CONFORMALITY IN BRACHYTHERAPY AND EXTERNAL-BEAM IRRADIATION - APPLICATION TO THE PROSTATE, International journal of radiation oncology, biology, physics, 37(3), 1997, pp. 731-736
Citations number
19
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
03603016
Volume
37
Issue
3
Year of publication
1997
Pages
731 - 736
Database
ISI
SICI code
0360-3016(1997)37:3<731:ACNTQT>2.0.ZU;2-2
Abstract
Purpose: This article presents a method of quantitative assessment of the degree of conformality and its designation by a single numerical v alue. Methods and Materials: A conformation number is introduced to ev aluate objectively the degree of conformality, A comparison is made be tween the conformation number as found for external beam treatment pla ns and ultrasonically guided I-125 Seed implants for localized prostat e cancer. Results: The conformation number in case of a planning targe t volume irradiated with two opposed open beams, three open beams, and three beams with customized blocks amounted to 0.17, 0.39, and 0.65, respectively, The conformation number as found for ultrasonically guid ed permanent prostate implants using I-125 seeds averaged 0.72. Conclu sions: The conformation number is a convenient instrument for indicati ng the degree of conformality by a single numerical value, Treatments with a conformation number greater than 0.60 might be termed conformal radiotherapy. (C) 1997 Elsevier Science Inc.