The articulating couple PE-PMMA as a 'wear-model'

Citation
G. Scheller et al., The articulating couple PE-PMMA as a 'wear-model', Z ORTHOP GR, 136(6), 1998, pp. 566-567
Citations number
2
Categorie Soggetti
Ortopedics, Rehabilitation & Sport Medicine
Journal title
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE
ISSN journal
00443220 → ACNP
Volume
136
Issue
6
Year of publication
1998
Pages
566 - 567
Database
ISI
SICI code
0044-3220(199811/12)136:6<566:TACPAA>2.0.ZU;2-O
Abstract
The situation, described in this casuistry, shows the fatal consequences of the huge amount of wear particles from the PE-PMMA-components of a total h ip prosthesis. The prosthesis had been implanted in Russia. It was made of a PE-ball fixed to a titanium stem and a socket, which had been molded intr aoperatively in situ. The combination of PMMA and PE in the bearing surface caused an extrem osteolysis in the proximal femur and the acetabulum and t he loosening of the cup and the stem within only a few months.