Chromium hexacyanoferrate based glucose biosensor

Authors
Citation
Ms. Lin et Wc. Shih, Chromium hexacyanoferrate based glucose biosensor, ANALYT CHIM, 381(2-3), 1999, pp. 183-189
Citations number
14
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
381
Issue
2-3
Year of publication
1999
Pages
183 - 189
Database
ISI
SICI code
0003-2670(19990216)381:2-3<183:CHBGB>2.0.ZU;2-Z
Abstract
Chromium hexacyanoferrate possesses an electrocatalytic property for the re duction of hydrogen peroxide at 0.0 V and this mixed-valence cluster can al so be utilized as a matrix to immobilize enzyme onto the interface of a glu cose biosensor. The electrochemical glucose biosensor possesses a linearity up to 4 mM (R=0.9994) with sensitivity of 3.793 mu A/mM per cm(2) The appa rent Michaelis-Menten constant is 29.9 mM (calculated from the reciprocal p lot) under a steady-state condition. The detection limit is 3.1x10(-6) M (S IN=3). There are two interferences observed at the detection of glucose suc h as ascorbic acid (1.48%) and cysteine (2.78%) in the presence of 5.5 mM g lucose. However, no other interference was observed from the addition of ot her interference such as uric acid, acetaminophen, tyrosine and galactose. (C) 1999 Elsevier Science B.V. All rights reserved.