Comparative evaluation of a commercial system for identification of gram-positive cocci

Citation
H. Von Baum et al., Comparative evaluation of a commercial system for identification of gram-positive cocci, EUR J CL M, 17(12), 1998, pp. 849-852
Citations number
13
Categorie Soggetti
Microbiology
Journal title
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES
ISSN journal
09349723 → ACNP
Volume
17
Issue
12
Year of publication
1998
Pages
849 - 852
Database
ISI
SICI code
0934-9723(199812)17:12<849:CEOACS>2.0.ZU;2-N
Abstract
The performance of a new commercial system for the identification of differ ent groups of gram-positive cocci [BBL Crystal Gram-Positive (GP) Identific ation System; Becton Dickinson Microbiology Systems, Germany] was evaluated in comparison with two currently used commercial systems, the API Staph an d the API Strep (bioMerieux Diagnostic, Germany). A total of 191 strains fr om seven different gram-positive genera comprising 32 different species wer e tested. For the BBL Crystal GP system, the correct identification rate wi thout additional tests was 89.5% at the species level and 97.9% at the genu s level. The findings suggest that the newly introduced BBL Crystal GP ID s ystem provides an accurate method for the identification of gram-positive c occi, with an overall rate of correct species identification of about 90%, similar to that of the established API systems. Its major advantage is the extended spectrum of taxa included in a single test panel in contrast to th e two different API test kits. Furthermore, the simplicity of use and the s afe and rapid handling in a closed system conveniently accommodate existing laboratory work-flow.