Low-jitter trigger system for pulse calibration and intercomparison of high-speed samplers

Authors
Citation
Ng. Paulter, Low-jitter trigger system for pulse calibration and intercomparison of high-speed samplers, IEEE INSTR, 47(3), 1998, pp. 606-608
Citations number
7
Categorie Soggetti
Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
ISSN journal
00189456 → ACNP
Volume
47
Issue
3
Year of publication
1998
Pages
606 - 608
Database
ISI
SICI code
0018-9456(199806)47:3<606:LTSFPC>2.0.ZU;2-3
Abstract
A low-jitter (<1 ps) trigger system for pulse-waveform-based calibration an d intercomparison of high-speed samplers is described. The system uses a co mmercially available pulse generator and pulse splitter/delay line.