Jf. Eloy et al., Pump and probe experimental setup with optoelectronically pulsed antenna for giga/terahertz spectroscopy, IEEE INSTR, 47(2), 1998, pp. 453-458
Citations number
17
Categorie Soggetti
Instrumentation & Measurement
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
A detailed description of a picosecond pump-probe optoelectronic pulsing (g
enerating electromagnetic transients in free space.) and sampling system is
presented. This system will be used for spectroscopy in a [10-100]-GHz ran
ge of materials such as semiconductors. The signal-to-noise ratio obtained
in this system is greater than 40 dB and the time resolution can be less th
an 500 fs, thereby permitting studies of very fast, low-amplitude transient
phenomena. For microdevices separated by 5 cm an E peak field amplitude of
80 V/m is detected.