"Severe" accumulation conditions in stripping voltammetry: Possibilities of their use

Citation
Yb. Kletenik et al., "Severe" accumulation conditions in stripping voltammetry: Possibilities of their use, J ANALYT CH, 54(1), 1999, pp. 44-48
Citations number
15
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF ANALYTICAL CHEMISTRY
ISSN journal
10619348 → ACNP
Volume
54
Issue
1
Year of publication
1999
Pages
44 - 48
Database
ISI
SICI code
1061-9348(199901)54:1<44:"ACISV>2.0.ZU;2-Q
Abstract
A procedure was developed for the accumulation of metals under "severe" con ditions of stripping voltammetry at solid renewal indicator electrodes. The effect of the parameters of severe accumulation (accumulation voltage V-ac , current density, and hydrodynamic characteristics) on the deposition rate of a number of metals is studied. It is shown that the use of high V-ac va lues sharply increases the rate of deposition of electroactive metals by Tw o to three orders of magnitude. It is also demonstrated that under severe c onditions of accumulation at solid renewal electrodes, the analytical signa l remains to be reproducible and depends linearly on the metal concentratio n and its accumulation time.