Validation of thin-layer chromatography quantitation determination with CCD camera and slit-scanning densitometer

Citation
M. Petrovic et al., Validation of thin-layer chromatography quantitation determination with CCD camera and slit-scanning densitometer, J AOAC INT, 82(1), 1999, pp. 25-30
Citations number
5
Categorie Soggetti
Agricultural Chemistry
Journal title
JOURNAL OF AOAC INTERNATIONAL
ISSN journal
10603271 → ACNP
Volume
82
Issue
1
Year of publication
1999
Pages
25 - 30
Database
ISI
SICI code
1060-3271(199901/02)82:1<25:VOTCQD>2.0.ZU;2-R
Abstract
Video densitometric and classical slit-scanning determinations of pesticide s on thin-layer chromatographic plates were validated for linearity, precis ion, and detection limit, A comparison of results showed that slit-scanning densitometry is more sensitive and more precise than video densitometry. H owever, according to validation requirements, relative standard deviation f or video densitometry is acceptable (3.5-5.3%). Linearities are very good a nd almost identical for both techniques. The main advantage of video techno logy is speed (a few seconds, compared with 20 min with scanning densitomet ry). Video technology also provides excellent archiving facility. Image and chromatographic data can be stored together, edited, and used for many tas ks.