Yy. Liu et al., Use of a fluorescence plate reader for measuring kinetic parameters with inner filter effect correction, ANALYT BIOC, 267(2), 1999, pp. 331-335
A general method is presented here for the determination of the K-m, k(cat)
, and k(cat)/K-m of fluorescence resonance energy transfer (FRET) substrate
s using a fluorescence plate reader. A simple empirical method for correcti
ng for the inner filter effect is shown to enable accurate and undistorted
measurements of these very important kinetic parameters, Inner filter effec
t corrected rates of hydrolysis of a FRET peptide substrate by hepatitis C
virus (HCV) NS3 protease at various substrate concentrations enabled measur
ement of a gm value of 4.4 +/- 0.3 mu M and k(cat)/K-m value of 96,500 +/-
5800 M-1 s(-1). These values are very close to the HPLC-determined K-m valu
e of 4.6 +/- 0.7 mu M and k(cat)/K-m value of 92,600 +/- 14,000 M-1 s(-1).
We demonstrate that the inner filter effect correction of microtiter plate
reader velocities enables rapid measurement of K-i and K-i' values and kine
tic inhibition mechanisms for HCV NS3 protease inhibitors. (C) 1999 Academi
c Press.