S. Mutlu et al., An immunosensor: Immobilization of anti-HBs antibody on glow-discharge treated piezoelectric quartz crystal for HBs-AG detection, ANAL LETTER, 32(2), 1999, pp. 317-334
In this study, an immunosensor with a piezoelectric quartz crystal was deve
loped. An oscillator system, in order to measure the frequency shifts cause
d by the increase of mass on the piezoelectric crystal, was used. The surfa
ces of the quartz crystals were treated by a glow-discharge technique utili
zing ethylene diamine monomer to substitute amino groups on the surface of
the crystal. These groups were activated by a bifunctional cross-linker, gl
utaraldehyde, to immobilize an antibody (so called anti-HBs). Optimum condi
tions for each of the modification steps for application on the quartz crys
tal surfaces were established and the results were represented for optimum
time, pH and concentration. The sensor was calibrated at optimum conditions
for HBs-Ag detection.