Simultaneous determination of estriol and estriol 3-sulfate in serum by column-switching semi-micro high-performance liquid chromatography with ultraviolet and electrochemical detection

Citation
N. Tagawa et al., Simultaneous determination of estriol and estriol 3-sulfate in serum by column-switching semi-micro high-performance liquid chromatography with ultraviolet and electrochemical detection, J CHROMAT B, 723(1-2), 1999, pp. 39-45
Citations number
33
Categorie Soggetti
Chemistry & Analysis
Journal title
JOURNAL OF CHROMATOGRAPHY B
ISSN journal
13872273 → ACNP
Volume
723
Issue
1-2
Year of publication
1999
Pages
39 - 45
Database
ISI
SICI code
1387-2273(19990219)723:1-2<39:SDOEAE>2.0.ZU;2-G
Abstract
A column-switching HPLC with semi-microcolumn enabled us a direct and simul taneous analysis of estriol (E3) and estriol 3-sulfate (E3 S) in human seru m in combination with ultraviolet (for E3 S) and electrochemical (for E3) d etectors. The mobile phases (phosphate buffer pH 7.0) contained 5 mM tetra- n-butylammonium ion (TBA) as a counter ion for E3 S. Serum samples were dil uted with 200 mM phosphate buffer (pH 7.0) containing 100 mM TBA, then inje cted to the pre-column. After serum proteins had flowed out from the pre-co lumn, E3 and E3 S were transferred to the enrichment column. Subsequently t he analytes were eluted to the analytical column. Detection limits of E3 an d E3 S in human serum were 2.5 ng/ml and 295 ng/ml. Serum E3 and E3 S level s (mean+/-SD) of umbilical artery from 18 full-term healthy neonates were 3 3+/-23 ng/ml and 1.26+/-0.59 mu g/ml, respectively. (C) 1999 Elsevier Scien ce B.V. All rights reserved.