A method for determining the information capacity of x-ray imaging scintillator detectors by means of luminescence and modulation transfer function measurements

Citation
I. Kandarakis et al., A method for determining the information capacity of x-ray imaging scintillator detectors by means of luminescence and modulation transfer function measurements, MED BIO E C, 37(1), 1999, pp. 25-30
Citations number
24
Categorie Soggetti
Multidisciplinary,"Instrumentation & Measurement
Journal title
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING
ISSN journal
01400118 → ACNP
Volume
37
Issue
1
Year of publication
1999
Pages
25 - 30
Database
ISI
SICI code
0140-0118(199901)37:1<25:AMFDTI>2.0.ZU;2-I
Abstract
A method to determine the information capacity of x-ray phosphor screens us ed in the detectors of medical imaging systems is described. Information ca pacity was determined via x-ray luminescence efficiency (XLE), modulation t ransfer function (MTF) and emission spectrum measurements. The method was a pplied to laboratory prepared screens from commonly employed phosphor mater ials. The screen coating weight varied from 50 mg cm(-2) to 140 mg cm(-2). Results indicated that information capacity decreased with screen coating t hickness but also depended on intrinsic phosphor properties (density, effec tive atomic number, intrinsic conversion efficiency, light wavelength). The Gd2O2S:Tb phosphor, exhibiting high density and effective atomic number, w as found to be superior to La2O2S:Tb and Y2O2S:Tb.