A. Ruiz-medina et al., Flow injection solid phase spectrofluorimetric determination of pyridoxinein presence of group B-vitamins, FRESEN J AN, 363(3), 1999, pp. 265-269
A flow-through optosensor has been prepared for the sensitive and selective
determination of pyridoxine (vitamin Bg) in aqueous solutions. The sensor
was developed in conjunction with a monochannel flow-injection analysis sys
tem with fluorimetric detection using Sephadex SP-C25 resin as an active so
rbent substrate. This method of determination is carried out without any de
rivatization. The wavelengths of excitation and emission were 295 and 385 n
m, respectively. When a HCl (10(-3) mol L-1)/NaCl (3 x 10(-2) mol L-1) solu
tion is used as carrier solution, the sensor responds linearly in the measu
ring range of 5-200, 10-400 and 50-1800 ng mL(-1) with detection limits of
0.33, 0.67, and 5.70 ng mL(-1) for 2000, 1000 and 200 mu L of sample volume
, respectively. The relative standard deviation for ten independent determi
nations is less than 0.75% for 0.2 and 1.0 mt of sample volumes used, and 1
.31% for 2.0 mi, of sample volume used. The method was satisfactorily appli
ed to the determination of vitamin Bg in pharmaceutical preparations.