Ks. Om et al., Static state hemodynamic variables estimation model for the moving-actuator type total artificial heart - Part I - Cardiac output estimation, INT J ARTIF, 22(1), 1999, pp. 21-26
Cardiac output estimation is a very important study for the artificial hear
t. In this paper, we developed a cardiac output estimation model for the mo
ving-actuator type total artificial heart (MA-TAH) that was developed at Se
oul National University Hospital. The proposed model is simple and provides
beat-by-beat mean cardiac output estimation. Moreover, it uses non-invasiv
ely acquired signals. Model parameters were adjusted with in vitro data by
least mean square (LMS) algorithm. Results showed that the proposed scheme
gives a mean estimation error of about 0.1 (l/min) for the given data. This
ensures the suitability of the proposed model.