Spectral response based calibration method of Tristimulus colorimeters

Authors
Citation
G. Eppeldauer, Spectral response based calibration method of Tristimulus colorimeters, J RES NAT I, 103(6), 1998, pp. 615-619
Citations number
24
Categorie Soggetti
Multidisciplinary,"Engineering Management /General
Journal title
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
ISSN journal
1044677X → ACNP
Volume
103
Issue
6
Year of publication
1998
Pages
615 - 619
Database
ISI
SICI code
1044-677X(199811/12)103:6<615:SRBCMO>2.0.ZU;2-Z
Abstract
A new method is described to calibrate tristimulus colorimeters for high ac curacy color measurements Instead of traditional lamp standards, modern, hi gh accuracy detector standards are suggested for calibration. After high ac curacy absolute spectral response determination of the tristimulus receiver s, color (spectral) correction and peak (amplitude) normalization can minim ize uncertainties caused by imperfect realizations of the Commission Intern ationale de l'Eclairage (CIE) color matching functions. As a result of the corrections, stable light sources of different spectral power distributions can be measured with an accuracy dominated by the sub tenths of a percent uncertainty of novel spectral response determinations.