Finite element analysis of a Gamma nail within a fractured femur

Citation
Cj. Wang et al., Finite element analysis of a Gamma nail within a fractured femur, MED ENG PHY, 20(9), 1998, pp. 677-683
Citations number
12
Categorie Soggetti
Multidisciplinary
Journal title
MEDICAL ENGINEERING & PHYSICS
ISSN journal
13504533 → ACNP
Volume
20
Issue
9
Year of publication
1998
Pages
677 - 683
Database
ISI
SICI code
1350-4533(199811/12)20:9<677:FEAOAG>2.0.ZU;2-8
Abstract
Failures of Gamma nails which treat unstable femoral fractures have been re ported. In this paper, a finite element model to include a Gamma nail withi n a fractured femur was used to investigate the stresses in the Gamma nail. The effects for different types of fracture were investigated. The results show that its use for subtrochanteric fractures will cause higher stresses at the lag screw and upper distal screw insertion holes in the nail than w hen used for femur neck fractures. (C) 1999 IPEM. Published by Elsevier Sci ence Ltd. All rights reserved.